Summary
Successful exploitation of this vulnerability could lead to a drop in the wireless capability of the Alaris PC Unit. In order to exploit this vulnerability, an attacker would need to gain access to the network associated with the affected devices and redirect the BD Alaris PC Unit 's authentication requests with a custom code and complete an authentication handshake based on the information extracted from the authentication requests. The Alaris PC Unit will continue to function as programmed; however, network-based services such as pre-populating the Alaris PC Unit with infusion parameters through EMR Interoperability or wirelessly updating the Alaris System Guardrails (DERS) will not be available.
Vulnerabilities 1
-
Unscored —
BD Alaris PC Unit, Model 8015, Versions 9.33.1 and earlier and BD Alaris Systems Manager, Versions 4.33 and earlier The…