Summary
Successful exploitation of these vulnerabilities may allow an attacker with physical access to the affected device to obtain patient protected health information (PHI), and/or compromise the integrity of the device. The affected device is not network connected and does not contain hardware to be network connected.
Vulnerabilities 5
-
Medium 6.9
Use of Password Hash with Insufficient Computational Effort for Boston Scientific Zoom Latitude
An attacker with physical access to Boston Scientific Zoom Latitude Model 3120 can remove the hard disk drive or create a specially crafted USB to extract the password hash for brute force reverse engineering of the system password.
-
Medium 6.5
Improper Access Control for Boston Scientific Zoom Latitude
A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world.
-
Medium 6.5
Missing Support Integrity Check for Boston Scientific Zoom Latitude
The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB.
-
Medium 6.5
Reliance on Component that is not Updateable for Boston Scientific Zoom Latitude
The affected device uses off-the-shelf software components that contain unpatched vulnerabilities. A malicious attacker with physical access to the affected device could exploit these vulnerabilities.
-
Medium 6.2
Missing Protection against Hardware Reverse Engineering Using Integrated Circuit Imaging Techniques for Boston Scientific Zoom Latitude
An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when…
Revision history
-
#1
ICSMA-21-273-01 Boston Scientific Zoom Latitude
References
-
ICS Advisory ICSMA-21-273-01 Web Version
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-21-273-01
-
Recommended Practices
https://us-cert.cisa.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf
-
Recommended Practices
https://us-cert.cisa.gov/ics/tips/ICS-TIP-12-146-01B