Summary
Successful exploitation of these vulnerabilities may allow an attacker unauthorized access to the IntelliBridge EC40 and80 Hub. This could allow an attacker to execute software, modify system configurations, or view/update files, including unidentifiable patient data.
Vulnerabilities 2
-
High 8.1
Philips IntelliBridge EC 40 and EC 80 Hub Use of Hard-coded Credentials
IntelliBridge EC 40 and 60 Hub (C.00.04 and prior) contains hard-coded credentials, such as a password or a cryptographic key, which it uses for its own inbound authentication, outbound communication to external components, or encryption of internal data.
-
High 8.1
Philips IntelliBridge EC 40 and EC 80 Hub Authentication Bypass Using an Alternate Path or Channel
The standard access path of the IntelliBridge EC 40 and 60 Hub (C.00.04 and prior) requires authentication, but the product has an alternate path or channel that does not require authentication.
Revision history
-
#1
ICSMA-21-322-01 Philips IntelliBridge EC 40 and EC 80 Hub
References
-
ICS Advisory ICSMA-21-322-01 Web Version
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-21-322-01
-
Recommended Practices
https://us-cert.cisa.gov/ics/alerts/ICS-ALERT-10-301-01
-
Recommended Practices
https://us-cert.cisa.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf
-
Recommended Practices
https://us-cert.cisa.gov/ics/tips/ICS-TIP-12-146-01B