Skip to content
OpenAdvisories
High ICSMA-22-251-01 4 CVEs

Baxter Sigma Spectrum Infusion Pump (Update A)

Published Sep 8, 2022 · Updated Sep 29, 2022 · Revision 2

Summary

Successful exploitation of these vulnerabilities could result in access to sensitive data and alteration of system configuration.

Vulnerabilities 4

  • Medium 5.5

    Unauthenticated network reconfiguration via TCP/UDP

    The Baxter Spectrum WBM does not perform mutual authentication with the gateway server host. This may allow an attacker to perform a man in the middle attack that modifies parameters making the network connection fail.

  • Medium 5.0

    Format String vulnerability

    The Baxter Spectrum WBM is susceptible to format string attacks via application messaging. An attacker could use this to read memory in the WBM to access sensitive information or cause a Denial of Service (DoS) on the WBM.

  • Medium 4.2

    Unencrypted internal storage of security credentials

    The Baxter Spectrum Wireless Battery Module (WBM) stores network credentials and PHI (only applicable to Spectrum IQ pumps using auto programming) in unencrypted form. An attacker with physical access to a device that hasn't had all data and settings erased m…

  • Low 3.1

    Format String vulnerability

    The Baxter Spectrum WBM (v16, v16D38) and Baxter Spectrum WBM (v17, v17D19, v20D29 to v20D32) when in superuser mode is susceptible to format string attacks via application messaging. An attacker could use this to read memory in the WBM to access sensitive in…

Revision history

  1. #2

    Baxter Sigma Spectrum Infusion Pump (Update A)

  2. #1

    Baxter Sigma Spectrum Infusion Pump (Update A)

References