Summary
Successful exploitation of these vulnerabilities could allow an attacker to take any action at the operating system level. A threat actor could impact settings, configurations, software, or data on the affected product; a threat actor could interact through the affected product via a connected network.
Vulnerabilities 2
-
Critical 10.0
CVE-2023-1968
Instruments with Illumina Universal Copy Service v2.x are vulnerable due to binding to an unrestricted IP address. An unauthenticated malicious actor could use UCS to listen on all IP addresses, including those capable of accepting remote communications.
-
High 7.4
CVE-2023-1966
Instruments with Illumina Universal Copy Service v1.x and v2.x contain an unnecessary privileges vulnerability. An unauthenticated malicious actor could upload and execute code remotely at the operating system level, which could allow an attacker to change se…