Skip to content
OpenAdvisories
Critical ICSMA-23-180-01 1 CVE

Medtronic Paceart Optima System

Published Jun 29, 2023 · Updated Jun 29, 2023

Summary

Successful exploitation of this vulnerability could result in remote code execution or a denial-of-service condition impacting a healthcare delivery organization's Paceart Optima system.

Vulnerabilities 1

  • Critical 9.8
    CVE-2023-31222 EPSS 28%

    Medtronic Paceart MSMQ Deserialization of Untrusted Data

    Deserialization of untrusted data in Microsoft Messaging Queuing Service in Medtronic's Paceart Optima versions 1.11 and earlier on Windows allows an unauthorized user to impact a healthcare delivery organization’s Paceart Optima system cardiac device causing…

Revision history

  1. #1

    Initial Publication

References