Summary
Successful exploitation of these vulnerabilities could allow an attacker to compromise sensitive data, hijack a session, modify firmware, make changes to system configurations, among other system impacts.
Vulnerabilities 8
-
Critical 9.8
XML External Entity attack in log4net
Apache log4net versions before 2.0.10 do not disable XML external entities when parsing log4net configuration files. This allows for XXE-based attacks in applications that accept attacker-controlled log4net configuration files.
-
High 8.2
Stored Cross-Site Scripting on User Import Functionality
A malicious file could be uploaded into a System Manager User Import Function resulting in a hijacked session.
-
Medium 6.9
Stored Cross-Site Scripting on Device Import Functionality
Alaris Systems Manager does not perform input validation during the Device Import Function.
-
Medium 6.8
PCU Configuration Lacks Authentication
The configuration from the PCU can be modified without authentication using physical connection to the PCU.
-
Medium 6.1
Lack of Cryptographic Security of IUI Bus
The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running.
-
Medium 5.2
Wireless Card Firmware Improperly Signed
The firmware update package for the wireless card is not properly signed and can be modified.
-
Low 3.5
CQI Data Sniffing
An insecure connection between Systems Manager and CQI Reporter application could expose infusion data to an attacker.
-
Low 3.0
Lack of Dataset Integrity Checking
A GRE dataset file within Systems Manager can be tampered with and distributed to PCUs.
Revision history
-
#2
Update A - Updates to Mitigations (BD has released BD Alaris System 12.3, which remediates CVE-2018-1285, CVE-2023-30563, CVE-2023-30564, and CVE-2023-30565 and partially remediates CVE-2023-30562)
-
#1
Initial Publication
References
-
ICSA Advisory ICSMA-23-194-01 - Web Version
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-23-194-01
-
Recommended Practices
https://www.cisa.gov/resources-tools/resources/ics-recommended-practices
-
Recommended Practices
https://www.cisa.gov/sites/default/files/publications/Cybersecurity_Best_Practices_for_Industrial_Control_Systems.pdf
-
Recommended Practices
https://www.cisa.gov/topics/industrial-control-systems
-
Recommended Practices
https://www.cisa.gov/uscert/sites/default/files/publications/emailscams0905.pdf
-
Recommended Practices
https://www.cisa.gov/uscert/ncas/tips/ST04-014
-
Recommended Practices
https://www.cisa.gov/uscert/ics/alerts/ICS-ALERT-10-301-01
-
Recommended Practices
https://us-cert.cisa.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf
-
Recommended Practices
https://www.cisa.gov/uscert/ics/tips/ICS-TIP-12-146-01B