Summary
Successful exploitation of these vulnerabilities could allow an attacker to replay the session of the logged in ISCV user and gain access to patient records.
Vulnerabilities 2
-
High 7.7
Philips Intellispace Cardiovascular (ISCV) Use of Weak Credentials
A token is created using the username, current date/time, and a fixed AES-128 encryption key, which is the same across all installations.
-
High 7.7
Philips Intellispace Cardiovascular (ISCV) Improper Authentication
A flaw exists in the Windows login flow where an AuthContext token can be exploited for replay attacks and authentication bypass.