Skip to content
OpenAdvisories
High ICSMA-25-072-01 2 CVEs

Philips Intellispace Cardiovascular (ISCV)

Published Mar 13, 2025 · Updated Mar 13, 2025

Summary

Successful exploitation of these vulnerabilities could allow an attacker to replay the session of the logged in ISCV user and gain access to patient records.

Vulnerabilities 2

  • High 7.7

    Philips Intellispace Cardiovascular (ISCV) Use of Weak Credentials

    A token is created using the username, current date/time, and a fixed AES-128 encryption key, which is the same across all installations.

  • High 7.7

    Philips Intellispace Cardiovascular (ISCV) Improper Authentication

    A flaw exists in the Windows login flow where an AuthContext token can be exploited for replay attacks and authentication bypass.

Revision history

  1. #1

    Initial Publication

References