Medium 4.8
Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Cleartext Transmission of Sensitive Information
Communications between Medtronic MiniMed MMT pumps and wireless accessories are transmitted in cleartext. A sufficiently skilled attacker could capture these transmissions and extract sensitive information, such as device serial numbers.
-
Medium 5.3
Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay
Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote c…
-