Product · Becton Dickinson & Co
BD Alarisâ„¢ Point-of-Care Unit (PCU) Model 8015
2 known vulnerabilities.
2 results · Clear filters · Feed
-
Medium 6.1
Lack of Cryptographic Security of IUI Bus
The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running.
-
Medium 6.8
PCU Configuration Lacks Authentication
The configuration from the PCU can be modified without authentication using physical connection to the PCU.