Skip to content
OpenAdvisories
High · 7.6 CVE-2014-5406 EUVD-2014-5294GHSA-vxf4-qh89-w2r6

Hospira LifeCare PCA Infusion System

Published Jul 6, 2015 · Updated Nov 3, 2025 · Assigned by icscert

Description

The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.

Affected products

VendorProductAffectedFixed
Hospira LifeCare PCA Infusion System ≤ 5.0 7.0

Advisories 1

References