Description
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Affected products
| Vendor | Product | Affected | Fixed |
|---|---|---|---|
| Hospira | Plum A+ Infusion System | <= 13.4 | — |
| Hospira | Plum A+3 Infusion System | <= 13.6 | — |
| Hospira | Symbiq Infusion System | <= 3.13 | — |
Advisories 1
- ICSA-15-161-01 Hospira Plum A+ and Symbiq Infusion Systems Vulnerabilities CISA ICS & IT advisories · Jun 6, 2025