Skip to content
OpenAdvisories
Unscored CVE-2015-3954 EUVD-2015-3985GHSA-x5v7-rmgq-cm8v

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, versio…

Published Mar 25, 2019 · Updated Aug 6, 2024 · Assigned by icscert

Description

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Affected products

VendorProductAffectedFixed
Hospira Plum A+ Infusion System <= 13.4 —
Hospira Plum A+3 Infusion System <= 13.6 —
Hospira Symbiq Infusion System <= 3.13 —

Advisories 1