Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" c…

Published Apr 25, 2018 · Updated Sep 17, 2024 · Assigned by icscert

Description

Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

Affected products

VendorProductAffectedFixed
Abbott Laboratories Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI. All versions of pacemakers manufactured prior to August 28, 2017 —