An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and …

Published Feb 13, 2017 · Updated Aug 5, 2024 · Assigned by icscert

Description

An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.

Affected products

VendorProductAffectedFixed
— St. Jude Merlin@home Transmitter before 8.2.2 St. Jude Merlin@home Transmitter before 8.2.2 —