Skip to content
OpenAdvisories
Unscored CVE-2018-19010 EUVD-2018-10728GHSA-6c7p-wgw2-hw4j

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. …

Published Jan 28, 2019 · Updated Sep 16, 2024 · Assigned by icscert

Description

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.

Affected products

VendorProductAffectedFixed
ICS-CERT Dräger Infinity Delta Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. —

Advisories 1

References