Description

Medtronic MiniMed Insulin Pumps

are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

Affected products

VendorProductAffectedFixed
Medtronic MiniMed 508 pump All versions —
Medtronic MiniMed Paradigm 511 pump All versions —
Medtronic MiniMed Paradigm 512/712 pumps All versions —
Medtronic MiniMed Paradigm 712E pump All versions —
Medtronic MiniMed Paradigm 515/715 pumps All versions —
Medtronic MiniMed Paradigm 522/722 pumps All versions —
Medtronic MiniMed Paradigm 522K/722K pumps All versions —
Medtronic MiniMed Paradigm 523/723 pumps ≤ Software Versions 2.4A —
Medtronic MiniMed Paradigm 523K/723K pumps ≤ Software Versions 2.4A —
Medtronic MiniMed Paradigm Veo 554/754 pumps ≤ Software Versions 2.6A —
Medtronic MiniMed Paradigm Veo 554CM/754CM pumps ≤ Software Versions 2.7A —

Advisories 1