Description
In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.
Affected products
| Vendor | Product | Affected | Fixed |
|---|---|---|---|
| Medtronic | Valleylab FT10 Energy Platform (VLFT10GEN) | ≤ 2.1.0, ≤ 2.0.3 | — |
| Medtronic | Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) | ≤ 1.20.2 | — |
Advisories 1
- ICSMA-19-311-01 Medtronic Valleylab FT10 and LS10 CISA ICS & IT advisories · Nov 7, 2019