Description
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.
Affected products
| Vendor | Product | Affected | Fixed |
|---|---|---|---|
| Dräger | SC 6002XL | SC 6002XL | — |
| Dräger | SC6802XL | SC6802XL | — |
| Dräger | SC 7000 | SC 7000 | — |
| Dräger | SC8000 | SC8000 | — |
| Dräger | SC90000 XL | SC90000 XL | — |