Skip to content
OpenAdvisories
High · 7.3 CVE-2020-10627 EUVD-2020-3075GHSA-r26f-v6j2-2h4j

Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an In…

Published Dec 1, 2021 · Updated Aug 4, 2024 · Assigned by icscert

Description

Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

Affected products

VendorProductAffectedFixed
Insulet Omnipod Insulin Management System 19191, 40160, ZXP425, ZXR425 —