Skip to content
OpenAdvisories
Unscored CVE-2020-12041 EUVD-2020-4357GHSA-gvrr-pwvw-5j3q

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on …

Published Jun 29, 2020 · Updated Aug 4, 2024 · Assigned by icscert

Description

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.

Affected products

VendorProductAffectedFixed
— Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum v6.x model 35700BAX, Baxter Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module v9,11,13,14,15,16,v20D29,v20D30,v20D31,v22D24, Baxter Spectrum v8.x with Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24,Baxter Spectrum Wireless Battery Module v17,v20D… —

Advisories 1

References