Description
SOOIL Developments Co Ltd DiabecareRS,AnyDana-i & AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i & AnyDana-A mobile apps doesn't use adequate measures to authenticate the communicating entities before exchanging keys, which allows unauthenticated, physically proximate attackers to eavesdrop the authentication sequence via Bluetooth Low Energy.
Affected products
| Vendor | Product | Affected | Fixed |
|---|---|---|---|
| — | SOOIL Developments CoLtd DiabecareRS,AnyDana-i,AnyDana-A | Dana DiabecareRS, AnyDana-i, AnyDana-A All versions prior to 3.0 | — |
Advisories 1
- ICSMA-21-012-01 SOOIL Dana Diabecare RS Products CISA ICS & IT advisories · Jan 12, 2021