Description
An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.
Advisories 1
- ICSMA-21-294-01 B. Braun Infusomat Space Large Volume Pump CISA ICS & IT advisories · Oct 21, 2021