Skip to content
OpenAdvisories
High · 8.6 CVE-2022-4992 EUVD-2022-55996GHSA-6jx8-hc3v-28hr

Dräger Infinity M540 VG4.1.1 Spoofed Network Message Handling DoS/Tampering

Published Jun 2, 2026 · Updated Jun 5, 2026 · Assigned by VulnCheck

Description

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.

Affected products

VendorProductAffectedFixed
Dräger Infinity Acute Care System < VG4.2, < VG4.1.1, < VG4.0.3 VG4.2
Dräger Standalone Infinity M540 patient monitor < VG4.2, < VG4.1.1 VG4.2