Description
Oxford Nanopore Technologies' MinKNOW software at or prior to version 24.11 creates a temporary file to store the local authentication token during startup, before copying it to its final location. This temporary file is created in a directory accessible to all users on the system. An unauthorized local user or process can exploit this behavior by placing a file lock on the temporary token file using the flock system call. This prevents MinKNOW from completing the token generation process. As a result, no valid local token is created, and the software is unable to execute commands on the sequencer. This leads to a denial-of-service (DoS) condition, blocking sequencing operations.
Affected products
| Vendor | Product | Affected | Fixed |
|---|---|---|---|
| Oxford Nano Technologies | MinKNOW | < 24.11 | 24.11 |
Advisories 1
- ICSMA-25-294-01 Oxford Nanopore Technologies MinKNOW CISA ICS & IT advisories · Oct 21, 2025