Skip to content
OpenAdvisories
High · 7.2 CVE-2025-14097 EUVD-2025-203895GHSA-rjfq-mp68-v4f7

Remote Code Execution Vulnerability in Radiometer Products

Published Dec 17, 2025 · Updated Dec 17, 2025 · Assigned by Radiometer

Description

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.

Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution:
Customers should ensure the following:
• The network is secure, and access follows best practices.
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                     
Exploit Status:

Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

Affected products

VendorProductAffectedFixed
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems Application software versions >= 3.5MR11 with Windows 10 operating system
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems Application software versions >= 6.20MR2 with Windows 10 operating system
Radiometer Medical Aps AQT90 FLEX Analyzers Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems Application software versions >= 8.13 MR2 with Windows 10 operating system

References