High · 8.3 CVE-2025-62575 EUVD-2025-200320GHSA-959v-frv4-7fj8

Mirion Medical EC2 Software NMIS BioDose Incorrect Permission Assignment for Critical Resource

Published Dec 2, 2025 · Updated Dec 2, 2025 · Assigned by icscert

Description

NMIS/BioDose V22.02 and previous versions rely on a Microsoft SQL Server database. The SQL user account 'nmdbuser' and other created accounts by default have the sysadmin role. This can lead to remote code execution through the use of certain built-in stored procedures.

Affected products

VendorProductAffectedFixed
Mirion Medical EC2 Software NMIS BioDose < 23.0 23.0