Skip to content
OpenAdvisories
High ICSMA-20-345-01 3 CVEs

Medtronic MyCareLink Smart

Published Dec 8, 2020 · Updated Dec 8, 2020

Summary

Successful exploitation of these vulnerabilities together could result in the attacker being able to modify or fabricate data from the implanted cardiac device being uploaded to the CareLink Network and remotely execute code on the MCL Smart Patient Reader device, which could allow control of a paired cardiac device. The exploitation must be initiated within Bluetooth signal proximity of the vulnerable product. Medtronic is currently unaware of any cyberattack, privacy breach, or patient harm as a result of these vulnerabilities.

Vulnerabilities 3

  • High 8.8

    Medtronic MyCareLink Smart Heap-based Buffer Overflow

    Medtronic MyCareLink Smart 25000 is  vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an a…

  • High 8.8

    Medtronic MyCareLink Smart Time-of-check Time-of-use Race Condition

    Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code …

  • High 8.0

    Medtronic MyCareLink Smart Improper Authentication

    Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an …

Revision history

  1. #1

    ICSMA-20-345-01 Medtronic MyCareLink

References