Skip to content
OpenAdvisories

Product · Medtronic

Smart Model 25000 Patient Reader

3 known vulnerabilities.

Smart Model 25000 Patient Reader at a glance

Last 12 months
0
Critical · all time
0
Exploited · all time
0
Public exploit · all time
0
New vulnerabilities per month · last 12 months
0 1 1 December 2025: 0 vulnerabilities Dec 2025 January 2026: 0 vulnerabilities February 2026: 0 vulnerabilities March 2026: 0 vulnerabilities Mar April 2026: 0 vulnerabilities May 2026: 0 vulnerabilities June 2026: 0 vulnerabilities Jun July 2026: 0 vulnerabilities August 2026: 0 vulnerabilities September 2026: 0 vulnerabilities Sep October 2026: 0 vulnerabilities November 2026: 0 vulnerabilities
Show as table
MonthVulnerabilitiesCritical
Nov 202600
Oct 202600
Sep 202600
Aug 202600
Jul 202600
Jun 202600
May 202600
Apr 202600
Mar 202600
Feb 202600
Jan 202600
Dec 202500
Nov 202500
Oct 202500
Sep 202500
Aug 202500
Jul 202500
Jun 202500
May 202500
Apr 202500
Mar 202500
Feb 202500
Jan 202500
Dec 202400

Latest advisories

  • High max 8.8

    Medtronic MyCareLink Smart

    Successful exploitation of these vulnerabilities together could result in the attacker being able to modify or fabricate data from the implanted cardiac device being uploaded to the CareLink Network and remotely execute code on the MCL Smart Patient Reader de…

3 results · Feed

  • High 8.8

    Medtronic MyCareLink Smart Time-of-check Time-of-use Race Condition

    Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code …

  • High 8.8

    Medtronic MyCareLink Smart Heap-based Buffer Overflow

    Medtronic MyCareLink Smart 25000 is  vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an a…

  • High 8.0

    Medtronic MyCareLink Smart Improper Authentication

    Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an …