  • High 8.8

    Medtronic MyCareLink Smart Time-of-check Time-of-use Race Condition

    Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code …

  • High 8.8

    Medtronic MyCareLink Smart Heap-based Buffer Overflow

    Medtronic MyCareLink Smart 25000 is  vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an a…

  • High 8.0

    Medtronic MyCareLink Smart Improper Authentication

    Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an …