Skip to content
OpenAdvisories

Product · Contec Health

CMS8000 Patient Monitor

4 known vulnerabilities.

CMS8000 Patient Monitor at a glance

Last 12 months
0
Critical · all time
1
Exploited · all time
0
Public exploit · all time
0
New vulnerabilities per month · last 12 months
0 1 1 December 2025: 0 vulnerabilities Dec 2025 January 2026: 0 vulnerabilities February 2026: 0 vulnerabilities March 2026: 0 vulnerabilities Mar April 2026: 0 vulnerabilities May 2026: 0 vulnerabilities June 2026: 0 vulnerabilities Jun July 2026: 0 vulnerabilities August 2026: 0 vulnerabilities September 2026: 0 vulnerabilities Sep October 2026: 0 vulnerabilities November 2026: 0 vulnerabilities
Show as table
MonthVulnerabilitiesCritical
Nov 202600
Oct 202600
Sep 202600
Aug 202600
Jul 202600
Jun 202600
May 202600
Apr 202600
Mar 202600
Feb 202600
Jan 202600
Dec 202500
Nov 202500
Oct 202500
Sep 202500
Aug 202500
Jul 202500
Jun 202500
May 202500
Apr 202500
Mar 202500
Feb 202510
Jan 202531
Dec 202400

Latest advisories

  • Critical max 9.8
    ICSMA-25-030-01 4 CVEs rev. 2

    Contec Health CMS8000 Patient Monitor (Update A)

    Successful exploitation of these vulnerabilities could allow an attacker to remotely send specially formatted UDP requests or connect to an unknown external network that would allow them to write arbitrary data, resulting in remote code execution. The device …

    first Jan 30, 2025

4 results · Feed

  • High 7.7

    The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP addres…

  • Medium 5.9

    Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor

    In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with th…

  • High 7.5

    Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor

    The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The func…

  • Critical 9.8

    Out-of-bounds Write vulnerability in Contec Health CMS8000 Patient Monitor

    Contec Health CMS8000 Patient Monitor is vulnerable to an out-of-bounds write, which could allow an attacker to send specially formatted UDP requests in order to write arbitrary data. This could result in remote code execution.