24952 MyCareLink Monitor

4 known vulnerabilities.

4 results · Clear filters · Feed

  • Medium 4.4

    Medtronic MyCareLink 24950 Patient Monitor Insufficient Verification of Data Authenticity

    Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload…

  • Medium 5.2

    Medtronic MyCareLink Patient Monitor Network Credential Weakness

    Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication.

  • Medium 6.2

    Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function

    Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physica…

  • Medium 6.4

    Medtronic MyCareLink Patient Monitor Use of Hard-coded Password

    Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to t…