Latest advisories
-
High max 7.1
Medtronic 2090 Carelink Programmer Vulnerabilities (Update C)
Successful exploitation of these vulnerabilities may allow an attacker with physical access to a 2090 Programmer to obtain per-product credentials to the software deployment network. These credentials grant access to the software deployment network, but acces…
first Feb 27, 2018 -
Medium max 4.6ICSMA-18-347-01 1 CVE
Medtronic 9790, 2090 CareLink, and 29901 Encore Programmers
As part of the intended functionality of this device, it may store protected health information (PHI) or personally identifiable information (PII). Successful exploitation of the vulnerability may allow an attacker with physical access to an affected programm…
4 results · Feed
-
Medium 4.6
Medtronic 9790, 2090 CareLink, and 29901 Encore Programmers Missing Encryption of Sensitive Data
Medtronic CareLink and Encore Programmers do not encrypt or do not sufficiently encrypt sensitive PII and PHI information while at rest .
-
High 7.1
Medtronic 2090 Carelink Programmer Improper Restriction of Communication Channel to Intended Endpoints
Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encaps…
-
Medium 4.9
Medtronic 2090 Carelink Programmer Storing Passwords in a Recoverable Format
Medtronic 2090 CareLink Programmer uses a per-product username and password that is stored in a recoverable format.
-
Medium 4.8
Medtronic 2090 Carelink Programmer Relative Path Traversal
Medtronic 2090 CareLink Programmer’s software deployment network contains a directory traversal vulnerability that could allow an attacker to read files on the system.