Description
The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information.
Affected products
| Vendor | Product | Affected | Fixed |
|---|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor | All | — |
Advisories 1
- ICSMA-22-244-01 Contec Health CMS8000 Patient Monitor (Update A) CISA ICS & IT advisories · Jan 30, 2025