Skip to content
OpenAdvisories

Product · Contec Health

CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor

5 known vulnerabilities.

CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor at a glance

Last 12 months
0
Critical · all time
0
Exploited · all time
0
Public exploit · all time
0
New vulnerabilities per month · last 12 months
0 1 1 December 2025: 0 vulnerabilities Dec 2025 January 2026: 0 vulnerabilities February 2026: 0 vulnerabilities March 2026: 0 vulnerabilities Mar April 2026: 0 vulnerabilities May 2026: 0 vulnerabilities June 2026: 0 vulnerabilities Jun July 2026: 0 vulnerabilities August 2026: 0 vulnerabilities September 2026: 0 vulnerabilities Sep October 2026: 0 vulnerabilities November 2026: 0 vulnerabilities
Show as table
MonthVulnerabilitiesCritical
Nov 202600
Oct 202600
Sep 202600
Aug 202600
Jul 202600
Jun 202600
May 202600
Apr 202600
Mar 202600
Feb 202600
Jan 202600
Dec 202500
Nov 202500
Oct 202500
Sep 202500
Aug 202500
Jul 202500
Jun 202500
May 202500
Apr 202500
Mar 202500
Feb 202500
Jan 202500
Dec 202400

Latest advisories

  • High max 7.5
    ICSMA-22-244-01 5 CVEs rev. 2

    Contec Health CMS8000 Patient Monitor (Update A)

    Successful exploitation of these vulnerabilities could allow a threat actor to cause a denial-of-service condition, modify firmware with physical access to the device, access a root shell, or employ hard-coded credentials to make configuration changes.

    first Sep 1, 2022

5 results · Feed

  • Low 3.0

    Contec Health CMS8000

    Multiple binary application files on the CMS8000 device are compiled with 'not stripped' and 'debug_info' compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify add…

  • Medium 5.7

    Contec Health CMS8000

    The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, t…

  • Medium 4.3

    Contec Health CMS8000

    Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient in…

  • High 7.5

    Contec Health CMS8000

    The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. …

  • Medium 6.8

    Contec Health CMS8000

    A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously m…