Skip to content
OpenAdvisories
High · 7.5 CVE-2022-38100 EUVD-2022-40702GHSA-w768-wqpr-555x

Contec Health CMS8000

Published Sep 13, 2022 · Updated Apr 16, 2025 · Assigned by icscert

Description

The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.

Affected products

VendorProductAffectedFixed
Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor All —

Advisories 1