Life2000 service tools for test and calibration do not support user authentication

Published Nov 14, 2024 · Updated Nov 15, 2024 · Assigned by Baxter

Description

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Affected products

VendorProductAffectedFixed
Baxter Life2000 Ventilation System 06.08.00.00 and prior —