8 known vulnerabilities.

Life2000 Ventilation System at a glance

Last 12 months
0
Critical · all time
8
Exploited · all time
0
Public exploit · all time
0
New vulnerabilities per month · last 12 months
0 1 1 December 2025: 0 vulnerabilities Dec 2025 January 2026: 0 vulnerabilities February 2026: 0 vulnerabilities March 2026: 0 vulnerabilities Mar April 2026: 0 vulnerabilities May 2026: 0 vulnerabilities June 2026: 0 vulnerabilities Jun July 2026: 0 vulnerabilities August 2026: 0 vulnerabilities September 2026: 0 vulnerabilities Sep October 2026: 0 vulnerabilities November 2026: 0 vulnerabilities
Show as table
MonthVulnerabilitiesCritical
Nov 202600
Oct 202600
Sep 202600
Aug 202600
Jul 202600
Jun 202600
May 202600
Apr 202600
Mar 202600
Feb 202600
Jan 202600
Dec 202500
Nov 202500
Oct 202500
Sep 202500
Aug 202500
Jul 202500
Jun 202500
May 202500
Apr 202500
Mar 202500
Feb 202500
Jan 202500
Dec 202400

Latest advisories

  • Critical max 10.0

    Baxter Life2000 Ventilation System

    Successful exploitation of these vulnerabilities could lead to information disclosure and/or disruption of the device's function without detection.

8 results · Feed

  • Critical 10.0

    Life2000 ventilator and Service PC lack sufficient audit logging capabilities

    The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unaut…

  • Critical 10.0

    Life2000 service tools for test and calibration do not support user authentication

    The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate…

  • Critical 9.3

    Life2000 Ventilator microcontroller lacks memory protection

    The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorize…

  • Critical 9.3

    Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates

    The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compro…

  • Critical 9.3

    Debug port on Life2000 Ventilator serial interface is enabled by default

    The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unint…

  • Critical 9.3

    Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator

    The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privi…

  • Critical 9.3

    No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator

    There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to devi…

  • Critical 9.3

    Improper data protection on Life2000 ventilator serial interface

    Improper data protection on the ventilator's serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.