Life2000 Ventilation System at a glance
- Last 12 months
- 0
- Critical · all time
- 8
- Exploited · all time
- 0
- Public exploit · all time
- 0
Show as table
| Month | Vulnerabilities | Critical |
|---|---|---|
| Nov 2026 | 0 | 0 |
| Oct 2026 | 0 | 0 |
| Sep 2026 | 0 | 0 |
| Aug 2026 | 0 | 0 |
| Jul 2026 | 0 | 0 |
| Jun 2026 | 0 | 0 |
| May 2026 | 0 | 0 |
| Apr 2026 | 0 | 0 |
| Mar 2026 | 0 | 0 |
| Feb 2026 | 0 | 0 |
| Jan 2026 | 0 | 0 |
| Dec 2025 | 0 | 0 |
| Nov 2025 | 0 | 0 |
| Oct 2025 | 0 | 0 |
| Sep 2025 | 0 | 0 |
| Aug 2025 | 0 | 0 |
| Jul 2025 | 0 | 0 |
| Jun 2025 | 0 | 0 |
| May 2025 | 0 | 0 |
| Apr 2025 | 0 | 0 |
| Mar 2025 | 0 | 0 |
| Feb 2025 | 0 | 0 |
| Jan 2025 | 0 | 0 |
| Dec 2024 | 0 | 0 |
Latest advisories
-
Critical max 10.0ICSMA-24-319-01 9 CVEs
Baxter Life2000 Ventilation System
Successful exploitation of these vulnerabilities could lead to information disclosure and/or disruption of the device's function without detection.
8 results · Feed
-
Critical 10.0
Life2000 ventilator and Service PC lack sufficient audit logging capabilities
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unaut…
-
Critical 10.0
Life2000 service tools for test and calibration do not support user authentication
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate…
-
Critical 9.3
Life2000 Ventilator microcontroller lacks memory protection
The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorize…
-
Critical 9.3
Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compro…
-
Critical 9.3
Debug port on Life2000 Ventilator serial interface is enabled by default
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unint…
-
Critical 9.3
Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privi…
-
Critical 9.3
No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to devi…
-
Critical 9.3
Improper data protection on Life2000 ventilator serial interface