Description
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.
Affected products
| Vendor | Product | Affected | Fixed |
|---|---|---|---|
| Baxter | Life2000 Ventilation System | 06.08.00.00 and prior | — |
Advisories 1
- ICSMA-24-319-01 Baxter Life2000 Ventilation System CISA ICS & IT advisories · Nov 14, 2024