Skip to content
OpenAdvisories
Critical · 9.3 CVE-2024-9832 EUVD-2024-50148GHSA-pggr-pwr5-55f8

No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator

Published Nov 14, 2024 · Updated Nov 18, 2024 · Assigned by Baxter

Description

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

Affected products

VendorProductAffectedFixed
Baxter Life2000 Ventilation System 06.08.00.00 and prior —

Advisories 1

References