Description
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Affected products
| Vendor | Product | Affected | Fixed |
|---|---|---|---|
| Baxter | Life2000 Ventilation System | 06.08.00.00 and prior | — |
Advisories 1
- ICSMA-24-319-01 Baxter Life2000 Ventilation System CISA ICS & IT advisories · Nov 14, 2024