Skip to content
OpenAdvisories
Critical · 9.3 CVE-2024-48973 EUVD-2024-43157GHSA-xx59-963p-2v9w

Debug port on Life2000 Ventilator serial interface is enabled by default

Published Nov 14, 2024 · Updated Nov 18, 2024 · Assigned by Baxter

Description

The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Affected products

VendorProductAffectedFixed
Baxter Life2000 Ventilation System 06.08.00.00 and prior —

Advisories 1