Description
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.
Affected products
| Vendor | Product | Affected | Fixed |
|---|---|---|---|
| Baxter | Life2000 Ventilation System | 06.08.00.00 and prior | — |
Advisories 1
- ICSMA-24-319-01 Baxter Life2000 Ventilation System CISA ICS & IT advisories · Nov 14, 2024