Summary
Successful exploitation of these vulnerabilities in system accessories could allow an attacker to gain access to sensitive information, modify settings or parameters, or perform arbitrary actions as an authenticated user.
Vulnerabilities 13
-
High 7.5
Fresenius Kabi Agilia Connect Infusion System uncontrolled resource consumption
Requests may be used to interrupt the normal operation of the device. When exploited, Fresenius Kabi Agilia Link+ version 3.0 must be rebooted via a hard reset triggered by pressing a button on the rack system.
-
High 7.3
Fresenius Kabi Agilia Connect Infusion System insufficiently protected credentials
The web application on Agilia Link+ version 3.0 implements authentication and session management mechanisms exclusively on the client-side and does not protect authentication attributes sufficiently.
-
High 7.3
Fresenius Kabi Agilia Connect Infusion System
Sensitive endpoints in Fresenius Kabi Agilia Link+ v3.0 and prior can be accessed without any authentication information such as the session cookie. An attacker can send requests to sensitive endpoints as an unauthenticated user to perform critical actions or…
-
High 7.3
Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm
Fresenius Kabi Agilia Link + version 3.0 does not enforce transport layer encryption. Therefore, transmitted data may be sent in cleartext. Transport layer encryption is offered on Port TCP/443, but the affected service does not perform an automated redirect …
-
High 7.3
Fresenius Kabi Agilia Connect Infusion System use of client side authentication
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 allows user input to be validated on the client side without authentication by the server. The server should not rely on the correctness of the data because users might not support o…
-
Medium 6.5
Fresenius Kabi Agilia Connect Infusion System plaintext storage of a password
An attacker with physical access to the host can extract the secrets from the registry and create valid JWT tokens for the Fresenius Kabi Vigilant MasterMed version 2.0.1.3 application and impersonate arbitrary users. An attacker could manipulate RabbitMQ que…
-
Medium 6.5
Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm
The SSL/TLS configuration of Fresenius Kabi Agilia Link + version 3.0 has serious deficiencies that may allow an attacker to compromise SSL/TLS sessions in different ways. An attacker may be able to eavesdrop on transferred data, manipulate data allegedly sec…
-
Medium 6.3
Fresenius Kabi Agilia Connect Infusion System hard coded credentials
Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 contains service credentials likely to be common across all instances. An attacker in possession of the password may gain privileges on all installations of this software.
-
Medium 5.9
Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 issues authentication tokens to authenticated users that are signed with a symmetric encryption key. An attacker in possession of the key can issue valid JWTs and impersonate arbitra…
-
Medium 5.4
Fresenius Kabi Agilia Connect Infusion System cross site scripting
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 is vulnerable to reflected cross-site scripting attacks. An attacker could inject JavaScript in a GET parameter of HTTP requests and perform unauthorized actions such as stealing int…
-
Medium 5.3
Fresenius Kabi Agilia Connect Infusion System exposure of information through directory listing
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 has the option for automated indexing (directory listing) activated. When accessing a directory, a web server delivers its entire content in HTML form. If an index file does not exis…
-
Medium 5.3
Fresenius Kabi Agilia Connect Infusion System files or directories accessible to external parties
Fresenius Kabi Agilia SP MC WiFi vD25 and prior has a default configuration page accessible without authentication. An attacker may use this functionality to change the exposed configuration values such as network settings.
-
Unscored —
A local file inclusion vulnerability in ExpertPDF 9.5.0 through 14.1.0 allows attackers to read the file contents from …
Revision history
-
#2
ICSMA-21-355-01 Fresenius Kabi Agilia Connect Infusion System (Update A)
-
#1
ICSMA-21-355-01 Fresenius Kabi Agilia Connect Infusion System
References
-
ICS Advisory ICSMA-21-355-01 Web Version
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-21-355-01
-
Recommended Practices
https://www.cisa.gov/uscert/sites/default/files/publications/emailscams_0905.pdf
-
Recommended Practices
https://www.cisa.gov/uscert/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf
-
Recommended Practices
https://www.cisa.gov/uscert/ics/tips/ICS-TIP-12-146-01B