Skip to content
OpenAdvisories

Vendor

Fresenius Kabi

12 known vulnerabilities.

Fresenius Kabi at a glance

Last 12 months
0
Critical · all time
0
Exploited · all time
0
Public exploit · all time
0
New vulnerabilities per month · last 12 months
0 1 1 December 2025: 0 vulnerabilities Dec 2025 January 2026: 0 vulnerabilities February 2026: 0 vulnerabilities March 2026: 0 vulnerabilities Mar April 2026: 0 vulnerabilities May 2026: 0 vulnerabilities June 2026: 0 vulnerabilities Jun July 2026: 0 vulnerabilities August 2026: 0 vulnerabilities September 2026: 0 vulnerabilities Sep October 2026: 0 vulnerabilities November 2026: 0 vulnerabilities
Show as table
MonthVulnerabilitiesCritical
Nov 202600
Oct 202600
Sep 202600
Aug 202600
Jul 202600
Jun 202600
May 202600
Apr 202600
Mar 202600
Feb 202600
Jan 202600
Dec 202500
Nov 202500
Oct 202500
Sep 202500
Aug 202500
Jul 202500
Jun 202500
May 202500
Apr 202500
Mar 202500
Feb 202500
Jan 202500
Dec 202400

Latest advisories

  • High max 7.5
    ICSMA-21-355-01 13 CVEs rev. 2

    Fresenius Kabi Agilia Connect Infusion System (Update A)

    Successful exploitation of these vulnerabilities in system accessories could allow an attacker to gain access to sensitive information, modify settings or parameters, or perform arbitrary actions as an authenticated user.

    first Dec 21, 2021

12 results · Feed

  • High 7.3

    Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm

    Fresenius Kabi Agilia Link + version 3.0 does not enforce transport layer encryption. Therefore, transmitted data may be sent in cleartext. Transport layer encryption is offered on Port TCP/443, but the affected service does not perform an automated redirect …

  • High 7.3

    Fresenius Kabi Agilia Connect Infusion System use of client side authentication

    Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 allows user input to be validated on the client side without authentication by the server. The server should not rely on the correctness of the data because users might not support o…

  • Medium 5.4

    Fresenius Kabi Agilia Connect Infusion System cross site scripting

    Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 is vulnerable to reflected cross-site scripting attacks. An attacker could inject JavaScript in a GET parameter of HTTP requests and perform unauthorized actions such as stealing int…

  • Medium 6.3

    Fresenius Kabi Agilia Connect Infusion System hard coded credentials

    Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 contains service credentials likely to be common across all instances. An attacker in possession of the password may gain privileges on all installations of this software.

  • Medium 5.3

    Fresenius Kabi Agilia Connect Infusion System files or directories accessible to external parties

    Fresenius Kabi Agilia SP MC WiFi vD25 and prior has a default configuration page accessible without authentication. An attacker may use this functionality to change the exposed configuration values such as network settings.

  • Medium 6.5

    Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm

    The SSL/TLS configuration of Fresenius Kabi Agilia Link + version 3.0 has serious deficiencies that may allow an attacker to compromise SSL/TLS sessions in different ways. An attacker may be able to eavesdrop on transferred data, manipulate data allegedly sec…

  • Medium 6.5

    Fresenius Kabi Agilia Connect Infusion System plaintext storage of a password

    An attacker with physical access to the host can extract the secrets from the registry and create valid JWT tokens for the Fresenius Kabi Vigilant MasterMed version 2.0.1.3 application and impersonate arbitrary users. An attacker could manipulate RabbitMQ que…

  • Medium 5.3

    Fresenius Kabi Agilia Connect Infusion System exposure of information through directory listing

    Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 has the option for automated indexing (directory listing) activated. When accessing a directory, a web server delivers its entire content in HTML form. If an index file does not exis…

  • Medium 5.9

    Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm

    Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 issues authentication tokens to authenticated users that are signed with a symmetric encryption key. An attacker in possession of the key can issue valid JWTs and impersonate arbitra…

  • High 7.3

    Fresenius Kabi Agilia Connect Infusion System

    Sensitive endpoints in Fresenius Kabi Agilia Link+ v3.0 and prior can be accessed without any authentication information such as the session cookie. An attacker can send requests to sensitive endpoints as an unauthenticated user to perform critical actions or…

  • High 7.3

    Fresenius Kabi Agilia Connect Infusion System insufficiently protected credentials

    The web application on Agilia Link+ version 3.0 implements authentication and session management mechanisms exclusively on the client-side and does not protect authentication attributes sufficiently.

  • High 7.5

    Fresenius Kabi Agilia Connect Infusion System uncontrolled resource consumption