High max 7.5
Fresenius Kabi Agilia Connect Infusion System (Update A)
Successful exploitation of these vulnerabilities in system accessories could allow an attacker to gain access to sensitive information, modify settings or parameters, or perform arbitrary actions as an authenticated user.
first Dec 21, 2021
5 results · Feed
-
High 7.3
Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm
Fresenius Kabi Agilia Link + version 3.0 does not enforce transport layer encryption. Therefore, transmitted data may be sent in cleartext. Transport layer encryption is offered on Port TCP/443, but the affected service does not perform an automated redirect …
-
Medium 6.5
Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm
The SSL/TLS configuration of Fresenius Kabi Agilia Link + version 3.0 has serious deficiencies that may allow an attacker to compromise SSL/TLS sessions in different ways. An attacker may be able to eavesdrop on transferred data, manipulate data allegedly sec…
-
High 7.3
Fresenius Kabi Agilia Connect Infusion System
Sensitive endpoints in Fresenius Kabi Agilia Link+ v3.0 and prior can be accessed without any authentication information such as the session cookie. An attacker can send requests to sensitive endpoints as an unauthenticated user to perform critical actions or…
-
High 7.3
Fresenius Kabi Agilia Connect Infusion System insufficiently protected credentials
The web application on Agilia Link+ version 3.0 implements authentication and session management mechanisms exclusively on the client-side and does not protect authentication attributes sufficiently.
-
High 7.5
Fresenius Kabi Agilia Connect Infusion System uncontrolled resource consumption
Requests may be used to interrupt the normal operation of the device. When exploited, Fresenius Kabi Agilia Link+ version 3.0 must be rebooted via a hard reset triggered by pressing a button on the rack system.