5 known vulnerabilities.

4 results · Clear filters · Feed

  • High 7.3

    Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm

    Fresenius Kabi Agilia Link + version 3.0 does not enforce transport layer encryption. Therefore, transmitted data may be sent in cleartext. Transport layer encryption is offered on Port TCP/443, but the affected service does not perform an automated redirect …

  • High 7.3

    Fresenius Kabi Agilia Connect Infusion System

    Sensitive endpoints in Fresenius Kabi Agilia Link+ v3.0 and prior can be accessed without any authentication information such as the session cookie. An attacker can send requests to sensitive endpoints as an unauthenticated user to perform critical actions or…

  • High 7.3

    Fresenius Kabi Agilia Connect Infusion System insufficiently protected credentials

    The web application on Agilia Link+ version 3.0 implements authentication and session management mechanisms exclusively on the client-side and does not protect authentication attributes sufficiently.

  • High 7.5

    Fresenius Kabi Agilia Connect Infusion System uncontrolled resource consumption

    Requests may be used to interrupt the normal operation of the device. When exploited, Fresenius Kabi Agilia Link+ version 3.0 must be rebooted via a hard reset triggered by pressing a button on the rack system.