Vendor
Contec Health
9 known vulnerabilities.
Products
3 results · Clear filters · Feed
-
High 7.7
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP addres…
-
High 7.5
Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The func…
-
High 7.5
Contec Health CMS8000
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. …