Critical 10.0
Life2000 ventilator and Service PC lack sufficient audit logging capabilities
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unaut…
-
Critical 10.0
Life2000 service tools for test and calibration do not support user authentication
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate…
-
Critical 9.3
Life2000 Ventilator microcontroller lacks memory protection
The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorize…
-
Critical 9.3
Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compro…
-
Critical 9.3
Debug port on Life2000 Ventilator serial interface is enabled by default
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unint…
-
Critical 9.3
Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privi…
-
Critical 9.3
No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to devi…
-
Critical 9.3
Improper data protection on Life2000 ventilator serial interface
Improper data protection on the ventilator's serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
-
Critical 10.0
Vulnerability in Baxter Connex Health Portal
In Connex health portal released before8/30/2024, SQL injection vulnerabilities were found that could have allowed an unauthenticated attacker to gain unauthorized access to Connex portal's database. An attacker could have submitted a crafted payload to Conn…
-
Critical 9.4
Vulnerability in Welch Allyn Configuration Tool Software
Insufficiently Protected Credentials vulnerability in Baxter Welch Allyn Configuration Tool may allow Remote Services with Stolen Credentials.This issue affects Welch Allyn Configuration Tool: versions 1.9.4.1 and prior.
-
Critical 9.1
Vulnerability in Baxter Welch Allyn Connex Spot Monitor
Use of Default Cryptographic Key vulnerability in Baxter Welch Allyn Connex Spot Monitor may allow Configuration/Environment Manipulation.This issue affects Welch Allyn Connex Spot Monitor in all versions prior to 1.52.